II. Indications
- Emergency stabilization of Unstable Bradycardia, where Transcutaneous Pacing is inadequate
- Unstable Bradycardia (e.g. third degree AV Block)
- Prolonged transport time with higher likelihood of rhythm decompensation
-
Myocardial Infarction with new Arrhythmia (LBBB, RBBB, Type II second degree AV Block, Third Degree AV Block)
- High risk of fatal Bradycardia (up to 43%)
- However, first priority is reperfusion!
III. Advantages: Contrast with Transcutaneous Pacing
- Requires only one tenth of the delivered energy (milliamps) as Transcutaneous Pacing
- Does not require nearly the same level of sedation and analgesia as Transcutaneous Pacing
- Does not generate the significant tracing artifacts seen with Transcutaneous Pacing
IV. Equipment
- Monitoring
- Cardiac monitor
- Pulse Oximeter
- Blood Pressure monitor
- Intravenous Access
- Central Venous Line Placement
- Ultrasound
- Sterile gloves and gown
- Skin antiseptics (e.g. Chlorhexidine)
- Suture Material
- Sterile dressing
- Central venous sheath introducer (may be included in Pacemaker kit)
- Central venous sheath introducers have 2 ports
- Pacer wire port with diaphragm
- Side port (to infuse medications and fluids)
- Any Central Line may be used that is large enough to pass the pacer wire and balloon
- Central venous sheath introducers have 2 ports
- Transvenous Pacemaker Kit
- Bipolar Pacemaker Wire with balloon tip (and syringe for balloon)
- Transvenous Pacemaker wires are typically 60-70 cm long (marked in 10 cm increments)
- Balloon tip allows the wire to float with Blood Flow
- Passes through the tricuspid valve and into the right ventricle
- Right ventricular apex is the preferred target (but any wall in the ventricle will suffice)
- Distal wire has 2 small leads separated by millimeters
- Proximal lead (positive)
- Distal lead (negative)
- Proximal connectors attach to Pacemaker generator specific connectors
- Positive Lead (proximal)
- Negative lead (distal)
- Pacer wire protective sterile sheath
- Pacer wire securing tape or dressing
- Bipolar Pacemaker Wire with balloon tip (and syringe for balloon)
- External pacer devices
- Pulse generator box
- Connecting cables, wires
V. Technique
- Preferred Central Line sites (most direct, least tortuous courses)
- Right internal Jugular Vein (preferred)
- Left subclavian vein
- Left internal Jugular Vein (similar wire course as with the left subclavian vein)
- Femoral veins may be used as an alternative access point
- Place Catheter 6 French sheath (in Pacemaker kit)
- Use 6 French instead of 9 French standard Cordis catheter
- Larger, 9 French leaks blood and fails to allow adequate Pacemaker wire control
- Precaution: Ensure venous placement (non-arterial)
- Cordis catheter is a large bore catheter with significant complications if placed arterially
- Place using seldinger technique, as with Central Line Placement
- Needle placement into vein
- Passage of guide wire through needle and removal of introducer needle
- Incision at insertion site and dilator insertion
- Thread the sheath with stylet over wire and into the vein
- Remove the guide wire
- Check that blood may be aspirated from side port
- Use 6 French instead of 9 French standard Cordis catheter
- Attach the plastic, accordion sheath
- Connect the sheath hub to the catheter
- Before placing the wire, estimate the insertion length from introducer to right heart
- Hold the wire using sterile technique above the patient's chest to estimate
- Pacemaker wire threaded through the sterile sheath and into the catheter
- Test wire balloon by inserting 1.5 cc air prior to insertion (then deflate)
- Correct amount of air should be used (specific syringe is typically in kit)
- Balloon air volume may be device specific
- Attach the pacer
- Thread the wire
- Deflate balloon (after testing as above)
- Pass wire via sterile sheath into 6 French catheter
- Wire inserted to the second mark (20 cm)
- Allows catheter to pass the end of the introducer sheath and allow for balloon inflation
- Balloon inflation is described below
- Infusing fluid via the catheter side port may facilitate floating the balloon into the right ventricle
- Advance Wire
- Turn on Pacemaker generator once tip has cleared introducer sheath
- Rate: 100 bpm to allow for early identification of capture (decreased to 60-80 after capture)
- Output set to maximum current (20-25 mA)
- Setting to maximal output (20-25 mA) allows for early detection of capture
- Decrease output once capture occurs (see below)
- Sensitivity
- Device safety feature suppresses output when exposed to intrinsic or external beats
- Set to asynchronous, lowest possible sensitivity (or off) when floating the wire
- May increase sensitivity once capture is achieved
- Reinflate balloon (by assistant)
- Inflate balloon with 1.5 cc air and turn stop-cock
- Balloon reinflated to allow floating of wire into the right atrium and right ventricle
- Advance the wire to the third mark (30 cm) and beyond as needed
- Use a smooth continuous motion
- If failure to capture after insertion of pacer wire to estimated length, withdraw wire and try again
- Deflate balloon
- Withdraw wire to 20 cm
- Reinflate balloon
- Advance wire
- Multiple reattempts may be needed (see troubleshooting below)
- Phrenic nerve pacing may occur during pacer wire advancement
- Results in severe Tachypnea
- If phrenic pacing occurs, deflate balloon and advance pacer wire
- EKG Monitor (not the EKG machine or alligator clip)
- Observe for electrical and mechanical capture
- Electrical tracing will show a Pacemaker spike followed by Wide QRS (LBBB appearance)
- Pacemaker spike may not be seen on some monitors
- Confirmation of capture and lead placement
- Heart Rate will increase from Bradycardia to pacer rate at capture (with wide complexes)
- Palpate pulse or auscultate heart sounds
- Oxygen Saturation waveform
- Bedside Cardiac Ultrasound (Subxiphoid Echocardiogram View)
- If transcutaneous pacer is also attached
- Transcutaneous pacer rate should be at 60 bpm (to differentiate from transvenous pacer)
- Once transvenous capture occurs, turn off the transcutaneous pacer
- Deflate balloon once capture is achieved
- Open stopcock and allow balloon to deflate spontaneously (syringe fills with air)
- After balloon deflation, advance pacer wire another 0.5 cm
- If capture lost, reinflate balloon and advance again
- Turn on Pacemaker generator once tip has cleared introducer sheath
- Troubleshooting problems passing through the tricuspid valve
- Short Stature patient (sharp angle between superior vena cava and tricuspid valve)
- Withdraw the introducer by small increments and reattempt
- Tricuspid regurgitation pushes balloon backwards into right atrium
- Consider threading without balloon inflated
- Wire threaded into pulmonary artery
- Deflate balloon immediately after pacer capture is achieved
- Failure to capture at estimated pacer wire distance (see above)
- Withdraw and reattempt (may require 5 repeat trials before wire passes through tricuspid, into RV)
- Deflate balloon, withdraw wire to 20 cm (2nd mark), reinflate balloon, advance wire
- Modifications
- Change the speed of pacer wire insertion (slower or faster)
- Pacer wire with predetermined curve by manufacturer
- Attempt to pass wire without reinflating balloon
- Pull back wire to 20 cm, rotate wire 90 or 180 degrees and reattempt insertion
- Wire without predetermined curve (straight wire)
- Pull wire completely out of sheath, maintaining sterile technique
- Coil pacer wire around 2 fingers to introduce curve
- Reinsert and attempt insertion techniques for wires with predetermined curve
- Maneuver requiring patient cooperation
- Patient placed in right decubitus position
- Wire advanced with balloon inflated (may ease passage through tricuspid valve)
- Patient takes a deep breath (increasing right ventricular inflow)
- Wire is advanced during deep breath
- Patient placed in right decubitus position
- Consider alternative access site other than internal Jugular Vein
- Opposite internal Jugular Vein
- Femoral vein
- Withdraw and reattempt (may require 5 repeat trials before wire passes through tricuspid, into RV)
- Short Stature patient (sharp angle between superior vena cava and tricuspid valve)
- Adjust Pacemaker generator output
- Decrease pacer output by 50% (e.g. 12.5 mA) and observe for 5-10 seconds to confirm continued pacing
- Decrease pacer output by another 50% (e.g. 6.25 mA) and observe again for 5-10 seconds
- Continue to decrease to 3 mA, then 1 mA
- Decrease pacer output until capture is lost (typically 0.3 mA or less)
- Increase again until capture occurs and to approximately 2.5 times lowest capture threshold
- Typical pacer output: 2 mA
- If pacer output required for capture >5 mA, pacer wire is unlikely to be at right ventricular wall
- Most often at tricuspid valve annulus
- May attempt repositioning wire, but in emergencies, may be left in place until definitive management
- Secure catheter and pacer
- Extend sheath to cover pacing catheter and tighten the associated valve (tightens diaphragm around wire)
- Sheath may be Sutured to skin via a side tab
- Tape wire to patient's skin
- Consider confirming lead placement with Bedside Ultrasound
VI. Safety
- Safe and effective (95% sucess rate) when performed in the emergency department for Symptomatic Bradycardia
VII. References
- Bessman in Roberts (2014) Roberts and Hedges Emergency Procedures, Elsevier, Philadelphia, p. 277-97
- Kwon and Warrington (2016) Crit Dec Emerg Med 30(9):10-11
- Orman and Bellezzo in Herbert (2016) EM:Rap 16(4): 8-9
- Sacchetti in Herbert (2017) EM:Rap 17(5): 1-2
- Tycynska (2025) Crit Dec Emerg Med 39(8): 4-15