http://www.fpnotebook.com/
Pacemaker
Aka: Pacemaker
- See Also
- Implantable Defibrillator (AICD)
- Cardiac Resynchronization Implantable Device
- Left Ventricular Assist Device (LVAD)
- Indications: Permanent Pacemaker in AV Block
- Class I Indications (helpful)
- Sick Sinus Syndrome with symptomatic Bradycardia
- Second Degree AV Block with symptomatic Bradycardia
- Third Degree AV Block with one associated condition
- Symptomatic Bradycardia
- Documented Asystole (3 seconds or greater)
- Catheter ablation of the AV junction
- Neuromuscular disorder with AV Block
- Myotonic muscular disorder
- Kearns-Sayre Syndrome
- Erb's Dystrophy (limb-girdle)
- Peroneal muscular atrophy
- Class IIa Indications (probably helpful)
- Sick Sinus Syndrome with rate <40 bpm
- Symptom association with Bradycardia unclear
- Asymptomatic third degree AV Block
- Asymptomatic Type II second degree AV Block
- Asymptomatic Type I AV Block at His level
- First degree AV Block and Pacemaker syndrome symptoms
- Class IIb Indications (Possibly helpful)
- Sick Sinus Syndrome with chronic awake rate <30 bpm
- Minimally symptomatic patients
- Marked First Degree AV Block (>0.3 seconds) with CHF
- Class III Indications (Not helpful, possibly harmful)
- Asymptomatic Bradycardia due to medication
- Asymptomatic First Degree AV Block
- Asymptomatic Type I AV Block limited to supra-His
- Transient AV Block secondary to resolving condition
- Drug toxicity
- Lyme Disease
- Description: Pacemaker Codes
- Position 1 (chamber paced)
- V - Ventricle
- A - Atrium
- D - Dual (A and V)
- O - None
- Position 2 (chamber sensed)
- V - Ventricle
- A - Atrium
- D - Dual (A and V)
- O - None
- Position 3 (response to sensing)
- V - Triggered
- I - Inhibited
- D - Dual (T and I)
- O - None
- Position 4 (programmable functions and rate modulation)
- P - Programmable rate and output
- M - Muti-programmability of rate, output, sensitivity
- C - Communicating via telemetry
- R - Rate modulation
- O - None
- Position 5 (anti-tachyarrhythmia)
- P - Pacing (anti-tachyarrhythmia)
- S - Shock
- D - Dual (P and S)
- O - None
- Types: Pacemaker Selection for Sinus Node Dysfunction
- No signs or future risks for impaired AV conduction
- Rate response: Rate-responsive atrial pacer (AAIR)
- No rate response: Atrial Pacemaker (AAI)
- Impaired AV Conduction and no AV synchrony needed
- Rate response: Rate-responsive dual chamber (DDDR)
- No rate response: Ventricular Pacemaker (DDD)
- Impaired AV Conduction and AV synchrony needed
- Tachyarrhythmia (e.g. PSVT)
- Rate response: Rate-responsive dual chamber (DDDR)
- No rate response: Dual chamber Pacemaker (DDD)
- No Tachyarrhythmia
- Rate response: Rate-responsive dual and mode switch
- No rate response: Dual chamber with mode switching
- Precautions
- Magnetic field exposure
- Pacemakers typically switch to asynchronous pacing at a set rate on exposure to magnetic field
- MRI scans are contraindicated
- Exception: Recent devices as of 2012 that have been designed MRI safe (with limitations)
- Cell phones should be held on the opposite side of body, away from Pacemaker
- Theoretical risk only
- Magnets may be helpful in some emergency settings (e.g. applied to AICD that is delivering inappropriate shocks)
- Battery life remaining in Pacemaker
- Pacers are interrogated to determine remaining battery life
- When battery life drops below ERI (Elective replacement indicator), Heart Rate will be be fixed at a manufacturer-set rate
- When battery life drops below EOL (End of life), Heart Rate will be fixed at a different fixed manufacturer-set rate
- External Defibrillation and cardioversion
- Electrical shock may theoretically damage Pacemaker
- Emergency Defibrillation may be performed without regard to Pacemaker
- Avoid applying elective cardioversion pads directly over Pacemaker
- Management: Troubleshooting a malfunctioning Pacemaker
- Pacemaker problems
- Dislodged Pacemaker lead
- Pacemaker syndrome
- Loss of atrial capture with only ventricle paced
- Presents with retrograde pulsations into the neck and secondary Heart Failure or Hypotension
- Runaway Pacemaker syndrome
- A damaged Pacemaker theoretically could sporadically increase paced rates at well above 100
- Would require emergent removal of the device
- Inadequate pacer energy
- Interrogate the Pacemaker
- Pacemaker programming head is placed directly over Pacemaker and push interrogate button
- Identify the patient's underlying rhythm
- Are the pacer spikes atrial or ventricular?
- Is the rhythm Pacemaker dependent?
- If uncertain, Pacemaker rate can be slowed to see if patient's rate also drops
- Can the Pacemaker sense the heart rhythm?
- Can the Pacemaker pace the heart?
- Are there other rhythm problems?
- Some Pacemakers can store abnormal rhythm events (e.g. VT runs) with a date-time stamp
- Is the Pacemaker programmed correctly?
- Evaluation measures
- Electrocardiogram
- Chest XRay
- May identify misplaced or damaged lead
- Data to have available when communicating with Cardiology about a patient with possible Pacemaker-related problem
- Is the device a Pacemaker or Defibrillator?
- How many wires to heart are present?
- When and where was the device implanted?
- Complications: Pacemaker infection
- Pacemakers must be removed for either early or late infections
- Early infection
- Presents with localized erythema, swelling, purulent discharge within weeks of Pacemaker placement
- Infections occur more often in Diabetes Mellitus, post-placement hematoma or with Defibrillator placement
- Late infection
- Presents with insidious, slowly developing infection; may only demonstrate an overlying Skin Erosion
- Infections occur more often after Pacemaker manipulations (with 1-3% risk with each manipulation)
- References
- Jones and Orman in Majoewsky (2012) EM:Rap 12(5): 4-6
- Bernstein (1987) Pacing Clin Electrophysiol 10:794-9
- Gregoratos (1998) J Am Coll Cardiol 31:1175-209
- Gregoratos (2005) Am Fam Physician 71:1563-70